Applications
- Used for drying, baking, wax treatment, heat treatment, sterilization, bacterial culture, and incubation.
- Used for culturing bacteria and other microorganisms.
- The twin-chamber configuration allows different incubation functions to be performed simultaneously.
Salient Features
- Twin-chamber model comprising Upper Chamber and Lower Chamber.
- Upper Chamber: Bacteriological Incubator.
- Lower Chamber: BOD Incubator.
- Both chambers can also be configured for BOD incubation as applicable to the model.
- Both inner chambers and exterior body are fabricated from Stainless Steel SS-304 grade.
- 75 mm insulation provided in both chambers; puff insulation for BOD chamber and high-density glass wool insulation for bacteriological chamber, as applicable.
- Both chambers are equipped with forced-air circulation blowers and heating elements for uniform temperature distribution.
- Quiet-running blower circulates air vertically downward and recirculates it throughout the chamber.
- Long-lasting U-shaped tubular heating elements provide efficient heating, long service life, and minimum heat loss.
- Refrigerated models are equipped with a hermetically sealed CFC-free compressor using R-134a gas, with evaporation coil and condenser.
- Electric motors are located at the rear and protected with safety covers.
- Heavy-duty stainless-steel locks and door hinges with spring-roller mechanism.
- Adjustable shelves for accommodating test samples and glassware of varying dimensions.
- Adjustable air ventilation provided on the sides of both chambers.
- Unit mounted on castor wheels for easy movement.
- Internal circuit wiring conforms to international norms.
- Aesthetic external finish and appearance.
- Supplied with pilot indicating lamps, 2 m power cord and power plug.
- Operating supply: 220/230 V, 50 Hz, single-phase AC.
- Supplied complete with IQ, OQ and DQ documentation.
Trays for Samples
- Trays in both chambers are manufactured from strong stainless-steel rods to withstand heavy test specimens.
- Test specimens should be distributed equally among the trays.
- Rod-type construction facilitates improved air circulation.
- Distance between trays: 15–20 cm.
Safety Features
- High-temperature safety cut-off thermostat.
- Electrical overload protection with 10 A fuse.
Pre-Installation Mandatory Requirements
- Maintain room temperature preferably at 25°C using air conditioning or provide a well-ventilated room with an exhaust fan.
- Ambient temperature should not exceed 30°C.
- Stabilized electrical supply of 230 V, 50 Hz, 20 A, AC must be available for reliable operation.
- Use of a servo-controlled voltage stabilizer is recommended.
- Warranty/guarantee may be void in case of non-compliance with the above installation requirements.